Xuzhou Huiya Imp&Exp Co., Ltd.
Products
Contact Us
  • Contact Person : Ms. Liu Jenny
  • Company Name : Xuzhou Huiya Imp&Exp Co., Ltd.
  • Tel : 0086-516-83998180
  • Fax : 86-516-83998180
  • Address : Jiangsu,XuZhou,Room 4-4034 Building45,EastHuanghe Road
  • Country/Region : China
  • Zip : 221005

Brand Meditech Infusion Set

Brand Meditech Infusion Set
Product Detailed
1. Three kinds of infusion modes: rate mode, time mode and body weight mode. 2. Applicable for all brands of standard infusion

Brand Meditech Infusion Set  Model BD8000

a) Audio-Visual alarm gives out and stop working automatically when the air detector detects there is air bubble in the pipe, in this way it can reduce the risk of causing air embolism.

b) Audio-Visual alarm gives out when the pressure sensor detects the pressure is beyond setting range.

c) Built-in double CPU monitor to ensure the infusion process safer and more reliable with wide infusion rate range from 1~2000ml/h.

d) With bolus (fast speed) dosage function.

Technique itemTechnique specifications:
Overall maximum size:125*140*205mm
Maximum weight:   <2.8kg
Power supply:  AC100-240V.50/60HZ
Max power consumption:25VA
Infusion rate range setting:  1-2000ml/h
Infusion increment: 0.1ml/h
Bolus infusion rate:   1000ml/h
Battery: Rechargeable lithium polymer battery 7.4V, 1600mA; working for 4 hours (10ml/h) after fully being charged
Accumulated infusion volume display:0.1~9999.9ml
Infusion accuracy:  ±3%( after being adjusted, up to ±2%)
OCCLU value: 39.2kPa-137.2kPa, adjustable, pressure shown on the graphic.
Rate mode:1-2000ml/h
Time mode:   Volume limit: 0.1~999.9ml     time: 1~2000 minutes
Alarm function:  over, OCCLU, low voltage, installation error, air, CPU communications error,  pump stuck, Low battery
Electromagnetic compatibility:Meet the requirements of IEC60601-1-2(2001 version)
Electrical safety: Meet the requirements of IEC60601-1:1998, GB9706.1-2007
KVO rate:    Adjustable range: 1ml/h~5.0ml/h
Preset infusion volume setting:0.1~9999.9ml/h (0 means the setting function of close pre-set volume)
Applicable infusion sets:    Meet all kinds of standard brands of infusion sets.
Operation condition:   Environment temperature: 5°C~+40°C, air pressure (86~106)kPa, relative humidity less than or equal to 80%             
Storage condition:    Environment temperature: -40°C~+55°C, air pressure (50~106)kPa, relative humidity less than or equal to 95%
Selectable modes:   1. Rate mode   2.  Time mode    3.body weight mode
Body weight mode:  weight: 0.1~300kg   drug limit: 0.1~999.9mg   volume limit: 0.1~999.9ml  Dose: 0.1~9999.9 (subject to unit, dosage, liquid, weight and infusion) unit: mg/kg/h, ug/kg/min
Panel instructions:   infusion rate, infusion preset volume, accumulated infusion volume, battery capacity, battery charge indication,  AC power supply connection indication, over, OCCUL. ALARM,  air, Bed No. ,  pressure value , flow rate